5 Block USER INTERFACE Device Functions Proline Promass 83 MODBUS RS485
50 Endress+Hauser
5.4.2 Function group MULTIPLEX
USER INTERFACE C → CONTROL CAA
→
MAIN LINE CCA
→
ADDITIONAL LINE CEA
→
INFORMATION LINE CGA → CONFIGURATION 260
→
MULTIPLEX 262
Function description
USER INTERFACE → INFORMATION LINE → MULTIPLEX
! Note!
If you select the BATCHING OPERATING KEYS in the function ASSIGN (2600), the multiplex display functionality is
not available in the information line.
ASSIGN
MODBUS register:
Data type:
Access:
2620
2542
Integer
read/write
For assigning a second reading to be displayed in the main line alternately (every
10 seconds) with the value defined in the ASSIGN function (2600).
Options: standard
0 = OFF
1 = MASS FLOW
2 = MASS FLOW IN %
3 = VOLUME FLOW
4 = VOLUME FLOW IN %
5 = CORRECTED VOLUME FLOW
6 = CORRECTED VOLUME FLOW IN %
7 = DENSITY
8 = REFERENCE DENSITY
9 = TEMPERATURE
10 = MASS FLOW BARGRAPH IN %
11 = VOLUME FLOW BARGRAPH IN %
12 = CORRECTED VOLUME FLOW BARGRAPH IN %
15 = ACTUAL CURRENT
18 = ACTUAL FREQUENCY
20 = TOTALIZER 1
21 = TOTALIZER 2
22 = TOTALIZER 3
23 = TAG NAME
24 = OPERATING/SYSTEM CONDITIONS
26 = DISPLAY FLOW DIRECTION
Factory setting: OFF
Advanced selection: with the optional SW package CONCENTRATION
61 = TARGET MASS FLOW
62 = % TARGET MASS FLOW
63 = TARGET VOLUME FLOW
64 = % TARGET VOLUME FLOW
65 = CORRECTED TARGET VOLUME FLOW
66 = CARRIER MASS FLOW
67 = % CARRIER MASS FLOW
68 = CARRIER VOLUME FLOW
69 = % CARRIER VOLUME FLOW
70 = CORRECTED CARRIER VOLUME FLOW
71 = % BLACK LIQUOR
72 = ° BAUME
74 = ° API
75 = ° PLATO
76 = ° BALLING
77 = ° BRIX
78 = OTHER (_ _ _ _ flexible concentration)
(continued on next page)