Device Functions Proline Promass 83 MODBUS RS485 10 Block SPECIAL FUNCTION
Endress+Hauser 165
10.3.10 Function group TUBE DAMPING FLUCTUATION
SPECIAL FUNCTION H
→
DENSITY FUNCTIONS HAA
↓
BATCHING FUNCTION HCA
↓
ADVANCED DIAGNOSTICS HEA
→
CONFIGURATION
740
↓
ACQUISITION 741
↓
MASS FLOW 742
↓
DENSITY 743
↓
REFERENCE DENSITY 744
↓
TEMPERATURE 745
↓
TUBE DAMPING 746
↓
ELECTRODYNAMIC
SENSORS
747
↓
OPERATING FREQUENCY
FLUCTUATION
748
↓
TUBE DAMPING
FLUCTUATION
749
Function description
SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → TUBE DAMP. FLUCT.
REFERENCE VALUE
MODBUS register:
Data type:
Access:
7490
See Note
Float
read
The reference value for the fluctuation of the tube damping appears on the
display.
User interface:
5-digit floating-point number
! Note for MODBUS!
Depends on the option selected in the function SELECT REFERENCE
CONDITION:
•Option = FACTORY → MODBUS register = 6747
•Option = USER → MODBUS register = 6749
ACTUAL VALUE
MODBUS register:
Data type:
Access:
7491
6751
Float
read
The measured fluctuation of the tube damping appears on the display.
User interface:
5-digit floating-point number
MINIMUM VALUE
MODBUS register:
Data type:
Access:
7492
6753
Float
read
The lowest value of the tube damping fluctuation since the saved values were last
reset appears on the display.
User interface:
5-digit floating-point number