Device Functions Proline Promass 83 MODBUS RS485 5 Block USER INTERFACE
Endress+Hauser 35
5.1.2 Function group UNLOCKING/LOCKING
USER INTERFACE C CONTROL CAA BASIC CONFIGURATION 200
UNLOCKING/LOCKING 202
Function description
USER INTERFACE CONTROL UNLOCKING/LOCKING
ACCESS CODE
MODBUS register:
Data type:
Access:
2020
2508
Float
read/write
! Note!
This function is only relevant for local operation and has no effect on access via
MODBUS RS485 communication.
All data of the measuring system are protected against inadvertent change.
Programming is disabled and the settings cannot be changed until a code is
entered in this function. If you press the
O / S keys in any function, the
measuring system automatically goes to this function and the prompt to enter the
code appears on the display (when programming is disabled).
You can enable programming by entering your personal code (Factory setting =
83, see function DEFINE PRIVATE CODE (2021)).
User input: max. 4-digit number: 0 to 9999
! Note!
Programming is disabled if you do not press a key within 60 seconds following
automatic return to the HOME position.
You can also disable programming in this function by entering any number
(other than the defined private code).
The Endress+Hauser service organization can be of assistance if you mislay
your personal code.
DEFINE PRIVATE
CODE
MODBUS register:
Data type:
Access:
2021
2510
Float
read/write
! Note!
This function is only relevant for local operation and has no effect on access via
MODBUS RS485 communication.
For specifying a personal code for enabling programming in the function ACCESS
CODE.
User input: 0 to 9999 (max. 4-digit number)
Factory setting: 83
! Note!
Programming is always enabled with the code “0”.
Programming has to be enabled before this code can be changed. When
programming is disabled this function is not available, thus preventing others
from accessing your personal code.
STATUS ACCESS
MODBUS register:
Data type:
Access:
2022
2512
Integer
read
Indicates whether access to the function matrix is currently possible (ACCESS
CUSTOMER) or whether configuration is locked (LOCKED).
Display:
0 = LOCKED (parameterization disabled)
1 = ACCESS CUSTOMER (parameterization possible)
ACCESS CODE
COUNTER
MODBUS register:
Data type:
Access:
2023
2568
Integer
read
Displays how often the customer code, service code or the digit “0” (code-free)
has been entered to gain access to the function matrix.
Display:max. 7-digit number: 0 to 9999999
Factory setting: 0