3 Block MEASURED VARIABLES Device Functions Proline Promass 83 MODBUS RS485
12 Endress+Hauser
3.1.2 Function group ADDITIONAL VALUES
MEASURED
VARIABLES
A MEASURING VALUES AAA MAIN VALUES 000
ADDITIONAL VALUES 002
Function description
MEASURED VARIABLES MEASURING VALUES ADDITIONAL VALUES
TARGET MASS
FLOW
MODBUS register:
Data type:
Access:
0020
2023
Float
read
! Note!
Function is not available unless one of the following was selected:
in the function DENSITY FUNCTION (7000), see Page 125:
% MASS / % VOLUME
FLEXIBLE and in the function MODE (7021), see Page 127, the selection %
MASS 2D or % MASS 3D
The currently measured mass flow of the target fluid appears on the display.
Target fluid = carried material (e.g. lime powder).
Display: 5-digit floating-point number, including unit and sign
% TARGET MASS
FLOW
MODBUS register:
Data type:
Access:
0021
2025
Float
read
! Note!
Function is not available unless one of the following was selected:
in the function DENSITY FUNCTION (7000), see Page 125:
% MASS / % VOLUME
FLEXIBLE and in the function MODE (7021), see Page 127, the selection %
MASS 2D or % MASS 3D
The currently measured mass flow of the target fluid is displayed as a % (of the
overall mass flow). Target fluid = carried material (e.g. lime powder).
Display: 5-digit floating-point number, including unit and sign
TARGET VOLUME
FLOW
MODBUS register:
Data type:
Access:
0022
2029
Float
read
! Note!
Function is not available unless one of the following was selected:
in the function DENSITY FUNCTION (7000), see Page 125:
% MASS / % VOLUME
FLEXIBLE and in the function MODE (7021), see Page 127, the selection %
VOLUME 2D or % VOLUME 3D
The currently measured volume flow of the target fluid appears on the display.
Target fluid = carried material (e.g. lime powder).
Display: 5-digit floating-point number, including unit and sign
% TARGET
VOLUME FLOW
MODBUS register:
Data type:
Access:
0023
2031
Float
read
! Note!
Function is not available unless one of the following was selected:
in the function DENSITY FUNCTION (7000), see Page 125:
% MASS / % VOLUME
FLEXIBLE and in the function MODE (7021), see Page 127, the selection %
VOLUME 2D or % VOLUME 3D
The currently measured volume flow of the target fluid is displayed as a % (of the
overall volume flow). Target fluid = carried material (e.g. lime powder).
Display: 5-digit floating-point number, including unit and sign
CORRECTED
TARGET VOLUME
FLOW
MODBUS register:
Data type:
Access:
0024
2033
Float
read
! Note!
Function is not available unless % MASS / % VOLUME was selected in the
DENSITY FUNCTION function (7000), (see Page 125).
The currently measured corrected volume flow of the target fluid appears on the
display.
Target fluid = carried material (e.g. lime powder).
Display: 5-digit floating-point number, including unit and sign