4 Block QUICK SETUP Device Functions Proline Promass 83 MODBUS RS485
30 Endress+Hauser
4.5 “Communication” Quick Setup menu
To establish serial data transfer, various arrangements between the MODBUS master and MODBUS
slave are required which have to be taken into consideration when configuring various functions.
These functions can be configured quickly and easily by means of the “Communication” Quick
Setup. The following table explains the parameter configuration options in more detail.
A0004430-en
Settings for the Communication Setup menu:
Fct. code Function name Suggested settings Description
Call up through the function matrix:
B QUICK SETUP QUICK SETUP COMMUNICATION see P. 20
1006 QUICK SETUP COMMUNICATION YES see P. 20
Basic configuration: Factory setting:
6301 FIELDBUS ADDRESS 247 see P. 107
6302 BAUDRATE 19200 BAUD see P. 107
6303 TRANSMISSION MODE RTU see P. 107
6304 PARITY EVEN see P. 107
6305 BYTE ORDER 1 - 0 - 3- 2 see P. 108
6306 DELAY TELEGRAM REPLY 10 ms see P. 108
+
+
E
+
E
B
Quick SetupQuick Setup
6301
6302
6306
1006
6303
6304
6305
Esc
E
+
-
XXX.XXX.XX
HOME-POSITION
Fieldbus
Address
Baudrate
Transmission
Mode
QS
Communication
Parity
Byte
Order
Delay
Telegram reply