10 Block SPECIAL FUNCTION Device Functions Proline Promass 83 MODBUS RS485
150 Endress+Hauser
10.3.2 Function group ACQUISITION
SPECIAL FUNCTION H DENSITY FUNCTIONS HAA
BATCHING FUNCTION HCA
ADVANCED DIAGNOSTICS HEA CONFIGURATION 740
ACQUISITION 741
Function description
SPECIAL FUNCTION ADVANCED DIAGNOSTICS ACQUISITION
ACQUISITION
MODE
MODBUS register:
Data type:
Access:
7410
6710
Integer
read/write
Use this function to specify whether the advanced diagnosis parameters should be
determined on a periodical or single-shot basis.
Options:
0 = OFF
1 = PERIODICAL
2 = SINGLE SHOT
Factory setting: OFF
! Note!
See the Chapter on “Commissioning” in the Operating Instructions BA107D for
more information on advanced diagnosis.
ACQUISITION
PERIOD
MODBUS register:
Data type:
Access:
7411
6501
Float
read/write
! Note!
Function is not available unless PERIODICAL was selected in the ACQUISITION
MODE function (7410).
Use this function to specify the time interval after which the advanced diagnosis
parameters should be taken. The time interval starts with the confirmation of the
input.
User input: 0 to 99999 s
Factory setting: 3600 s
! Note!
A reference status must be defined prior to determining the diagnosis parameters,
see function SELECT REFERENCE CONDITION (7402).
ACQUISITION DO
MODBUS register:
Data type:
Access:
7412
6711
Integer
read/write
! Note!
Function is not available unless SINGLE SHOT was selected in the ACQUISITION
MODE function (7410).
Use this function to start determining the advanced diagnosis parameters on a
single-shot basis.
User input:
0 = CANCEL
1 = START
Factory setting: CANCEL
! Note!
A reference status must be defined prior to determining the diagnosis parameters,
see function SELECT REFERENCE CONDITION (7402).
RESET HISTORY
MODBUS register:
Data type:
Access:
7413
6712
Integer
read/write
Use this function to delete all history values.
User input:
0 = NO
1 = YES
Factory setting: NO