6 Block TOTALIZER Device Functions Proline Promass 83 MODBUS RS485
54 Endress+Hauser
6.1 Group TOTALIZER (1 to 3)
6.1.1 Function group CONFIGURATION
TOTALIZER D → TOTALIZER 1 DAA → CONFIGURATION 300
→
TOTALIZER 2 DAB → CONFIGURATION 300
→
TOTALIZER 3 DAC → CONFIGURATION 300
Function description
TOTALIZER → TOTALIZER (1 to 3) → CONFIGURATION
! Note!
The function descriptions below apply to totalizers 1 to 3; the totalizers are independently configurable.
ASSIGN
MODBUS register:
Totalizer 1
Totalizer 2
Totalizer 3
Data type:
Access:
3000
2601
2801
3001
Integer
read/write
For assigning a measured variable to the totalizer in question.
Options: standard
0 = OFF
1 = MASS FLOW
2 = VOLUME FLOW
3 = CORRECTED VOLUME FLOW
Factory setting: MASS FLOW
Advanced selection: with the optional SW package CONCENTRATION
4 = TARGET MASS FLOW
5 = TARGET VOLUME FLOW
6 = TARGET CORRECTED VOLUME FLOW
7 = CARRIER MASS FLOW
8 = CARRIER VOLUME FLOW
9 = CARRIER CORRECTED VOLUME FLOW
! Note!
• The totalizer is reset to “0” as soon as the selection is changed.
• If you select OFF in the function group CONFIGURATION of the totalizer in
question, only the ASSIGN(3000) function remains visible.
UNIT TOTALIZER
MODBUS register:
• Totalizer 1
–Mass flow
– Volume flow
– Corr. vol. flow
• Totalizer 2
–Mass flow
– Volume flow
– Corr. vol. flow
• Totalizer 3
–Mass flow
– Volume flow
– Corr. vol. flow
Data type:
Access:
3001
2602
2603
2604
2802
2803
2804
3002
3003
3004
Integer
read/write
For selecting the unit for the measured variable assigned in the function ASSIGN
(3000).
Options: for the MASS FLOW assignment
0 to 2 = metric → g; kg; t
3 to 4 = US → oz; lb; ton
Factory setting: kg
Options: for the VOLUME FLOW assignment
0 to 6 = metric → cm
3
; dm
3
; m
3
; ml; l; hl; Ml Mega
7 to 16 = US → cc; af; ft
3
; oz f; gal; Mgal; bbl (normal fluids); bbl (beer);
bbl (petrochemicals); bbl (filling tanks)
22 = Kgal
17 to 20 = Imperial → gal; Mgal; bbl (beer); bbl (petrochemicals)
Factory setting: m
3
Options: for the CORRECTED VOLUME FLOW assignment
0 to 1 = metric → Nl; Nm
3
2 to 3 = US → Sm
3
; Scf
Factory setting: Nm
3