Device Functions Proline Promass 83 MODBUS RS485 6 Block TOTALIZER
Endress+Hauser 57
6.2 Group HANDLING TOTALIZER
TOTALIZER D → TOTALIZER 1 DAA
→
TOTALIZER 2 DAB
→
TOTALIZER 3 DAC
→
HANDLING TOTALIZER DJA → Handling totalizer functions
Function description
TOTALIZER → HANDLING TOTALIZER → Handling totalizer functions
RESET ALL
TOTALIZERS
MODBUS register:
Data type:
Access:
3800
2609
Integer
read/write
Resets the totals and the overflows of all totalizers to zero.
Options:
0 = NO
1 = YES
Factory setting: NO
! Note!
If the device has a status input and if it is appropriately configured, a reset for the
totalizer (1 to 3) can also be triggered by a pulse (see the ASSIGN STATUS INPUT
function (5000) on Page 103).
FAILSAFE ALL
TOTALIZERS
MODBUS register:
Data type:
Access:
3801
2607
Integer
read/write
Use this function to define the common response of all totalizers (1 to 3) to error.
Options:
0 = STOP
The totalizer is paused until the fault is rectified.
1 = ACTUAL VALUE
The totalizer continues to count based on the current flow measuring value. The
fault is ignored.
2 = HOLD VALUE
The totalizer continues to count the flow that is based on the last valid flow value
(before the fault occurred).
Factory setting: STOP