Device Functions Proline Promass 83 MODBUS RS485 10 Block SPECIAL FUNCTION
Endress+Hauser 159
10.3.7 Function group TUBE DAMPING
SPECIAL FUNCTION H → DENSITY FUNCTIONS HAA
→
BATCHING FUNCTION HCA
→
ADVANCED DIAGNOSTICS HEA → CONFIGURATION 740
→
ACQUISITION 741
→
MASS FLOW 742
→
DENSITY 743
→
REFERENCE DENSITY 744
→
TEMPERATURE 745
→
TUBE DAMPING 746
Function description
SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → TUBE DAMPING
REFERENCE VALUE
MODBUS register:
Data type:
Access:
7460
See Note
Float
read
The reference value for tube damping appears on the display.
Display: 5-digit floating-point number
! Note for MODBUS!
Depends on the option selected in the function SELECT REFERENCE
CONDITION:
•Option = FACTORY → MODBUS register = 6639
•Option = USER → MODBUS register = 6641
ACTUAL VALUE
MODBUS register:
Data type:
Access:
7461
6643
Float
read
The measured tube damping appears on the display.
Display: 5-digit floating-point number
MINIMUM VALUE
MODBUS register:
Data type:
Access:
7462
6645
Float
read
The lowest tube damping value since the saved values were last reset appears on
the display.
Display: 5-digit floating-point number
MAXIMUM VALUE
MODBUS register:
Data type:
Access:
7463
6647
Float
read
The highest tube damping value since the saved values were last reset appears on
the display.
Display: 5-digit floating-point number