10 Block SPECIAL FUNCTION Device Functions Proline Promass 83 MODBUS RS485
160 Endress+Hauser
HISTORY TUBE
DAMPING
MODBUS register:
Data type:
Access:
7464
See Note
Float
read
The last ten tube damping values since the saved values were last reset appears on
the display.
Display: 5-digit floating-point number
! Note for MODBUS!
The various histories are available via the following MOBBUS registers:
• History 1 = MODBUS register 6649
• History 2 = MODBUS register 6651
• History 3 = MODBUS register 6653
• History 4 = MODBUS register 6655
• History 5 = MODBUS register 6657
• History 6 = MODBUS register 6659
• History 7 = MODBUS register 6661
• History 8 = MODBUS register 6663
• History 9 = MODBUS register 6665
• History 10 = MODBUS register 6667
ACTUAL
DEVIATION
MODBUS register:
Data type:
Access:
7465
6671
Float
read
This function displays the deviation between the measured tube damping and the
reference values (FACTORY or USER), see Page 149, selected in the function
SELECT REFERENCE CONDITION (7402).
Display: 5-digit floating-point number
WARNING LEVEL
MODBUS register:
Data type:
Access:
7466
6669
Float
read/write
! Note!
Function is not available unless ON was selected in the WARNING MODE
function (7403).
Use this function to specify a limit value for tube damping. A notice message is
generated if the limit value is exceeded.
User input: 0 to 99999 [%]
Factory setting: 1000%
Function description
SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → TUBE DAMPING