Device Functions Proline Promass 83 MODBUS RS485 5 Block USER INTERFACE
Endress+Hauser 39
5.2.2 Function group MULTIPLEX
USER INTERFACE C → CONTROL CAA
→
MAIN LINE CCA → CONFIGURATION 220
→
MULTIPLEX 222
Function description
USER INTERFACE → CONTROL → MULTIPLEX
ASSIGN
MODBUS register:
Data type:
Access:
2220
2522
Integer
read/write
For assigning a second reading to be displayed in the main line alternately (every
10 seconds) with the value defined in the ASSIGN function (2200).
Options: (standard)
0 = OFF
1 = MASS FLOW
2 = MASS FLOW IN %
3 = VOLUME FLOW
4 = VOLUME FLOW IN %
5 = CORRECTED VOLUME FLOW
6 = CORRECTED VOLUME FLOW IN %
7 = DENSITY
8 = REFERENCE DENSITY
9 = TEMPERATURE
15 = ACTUAL CURRENT
18 = ACTUAL FREQUENCY
20 = TOTALIZER 1
21 = TOTALIZER 2
22 = TOTALIZER 3
Factory setting: OFF
Advanced selection: (with the optional SW package BATCHING)
52 = BATCH NAME (“BATCH # 1” or “BEER 330”, etc.)
53 = BATCH QUANTITY (overall quantity to be batched)
54 = BATCH COUNTER (batching processes carried out)
55 = BATCH SUM (effective total batching quantity)
! Note!
The options given in the BATCHING software package always refer to the
batching selected in the function BATCH SELECTOR (“BATCH # 1”,
“BATCH # 2”, etc.). Example: If BATCH # 1 was selected in the BATCH
SELECTOR function (7200), then only the values from BATCH # 1 (batch name,
batch quantity etc.) can be displayed.
Advanced selection: (with the optional SW package CONCENTRATION)
61 = TARGET MASS FLOW
62 = % TARGET MASS FLOW
63 = TARGET VOLUME FLOW
64 = % TARGET VOLUME FLOW
65 = CORRECTED TARGET VOLUME FLOW
66 = CARRIER MASS FLOW
67 = % CARRIER MASS FLOW
68 = CARRIER VOLUME FLOW
69 = % CARRIER VOLUME FLOW
70 = CORRECTED CARRIER VOLUME FLOW
71 = % BLACK LIQUOR
72 = ° BAUME
74 = ° API
75 = ° PLATO
76 = ° BALLING
77 = ° BRIX
78 = OTHER (_ _ _ _ flexible concentration)
(continued on next page)