10 Block SPECIAL FUNCTION Device Functions Proline Promass 83 MODBUS RS485
154 Endress+Hauser
ACTUAL
DEVIATION
MODBUS register:
Data type:
Access:
7435
6569
Float
read
This function displays the deviation between the measured density and the
reference values (FACTORY or USER), see Page 149, selected in the function
SELECT REFERENCE CONDITION (7402).
Display: 5-digit floating-point number, including units
WARNING LEVEL
MODBUS register:
Data type:
Access:
7436
6567
Float
read/write
! Note!
Function is not available unless ON was selected in the WARNING MODE
function (7403).
Use this function to specify a limit value for the density. A notice message is
generated if the limit value is exceeded.
User input: 0 to 99999 [%]
Factory setting: 100%
Function description
SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → DENSITY