10 Block SPECIAL FUNCTION Device Functions Proline Promass 83 MODBUS RS485
162 Endress+Hauser
HISTORY ELECTR.
SENSORS
MODBUS register:
Data type:
Access:
7474
See Note
Float
read
The last ten values of the electrodynamic sensors since the saved values were last
reset appear on the display.
Display: 5-digit floating-point number
! Note for MODBUS!
The various histories are available via the following MOBBUS registers:
• History 1 = MODBUS register 6683
• History 2 = MODBUS register 6685
• History 3 = MODBUS register 6687
• History 4 = MODBUS register 6689
• History 5 = MODBUS register 6691
• History 6 = MODBUS register 6693
• History 7 = MODBUS register 6695
• History 8 = MODBUS register 6697
• History 9 = MODBUS register 6699
• History 10 = MODBUS register 6701
ACTUAL
DEVIATION
MODBUS register:
Data type:
Access:
7475
6705
Float
read
This function displays the deviation between the measuring values for the
electrodynamic sensors and the reference values (FACTORY or USER), see
Page 149, selected in the function SELECT REFERENCE CONDITION (7402) is
displayed.
Display: 5-digit floating-point number
WARNING LEVEL
MODBUS register:
Data type:
Access:
7476
6703
Float
read/write
! Note!
Function is not available unless ON was selected in the WARNING MODE
function (7403).
Use this function to specify a limit value for the electrodynamic sensors. A notice
message is generated if the limit value is exceeded.
User input: 0 to 99999 [%]
Factory setting: 100%
Function description
SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → ELECTRODYNAMIC SENSORS