Device Functions Proline Promass 83 MODBUS RS485 4 Block QUICK SETUP
Endress+Hauser 21
T-DAT SAVE/LOAD
MODBUS register:
Data type:
Access:
1009
2401
Integer
read/write
Use this function to save the parameter settings / configuration of the transmitter
in a transmitter DAT (T-DAT), or to load the parameter settings from the T-DAT
into the EEPROM (manual safety function).
Application examples:
After commissioning, the current measuring point parameters can be saved to
the T-DAT as a backup.
If the transmitter is replaced for some reason, the data from the T-DAT can be
loaded into the new transmitter (EEPROM).
Options:
0 = CANCEL
1 = SAVE (from EEPROM to T-DAT)
2 = LOAD (from the T-DAT into EEPROM)
Factory setting:
CANCEL
! Note!
If the target device has an older software version, the message
“TRANSM. SW-DAT” is displayed during startup. Then only the SAVE function
is available.
•LOAD
This function is only possible if the target device has the same software version
as, or a more recent software version than, the source device.
•SAVE
This function is always available.
Function description