7 Block OUTPUTS Device Functions Proline Promass 83 MODBUS RS485
96 Endress+Hauser
MEASURING
MODE
MODBUS register:
Relay output 1
Relay output 2
Data type:
Access:
4705
3810
4010
Integer
read/write
! Note!
This function is not visible unless a limit value was assigned to the relay output.
Use this function to define the measuring mode for the relay output.
Options:
0 = STANDARD
The relay output signal switches at the defined switch points.
1 = SYMMETRY
The relay output signal switches at the defined switch points, irrespective of the
sign. If you define a switch point with a positive sign, the relay output switches as
soon as the value is reached in the negative direction (negative sign), (see
illustration).
Factory setting: STANDARD
A0001247
Fig. 31 : Example for the SYMMETRY measuring mode
Switch-on point Q = 4
Switch-off point Q = 10
m = Relay energized
n = Relay de-energized
! Note!
SYMMETRY cannot be selected unless the values in the ON-VALUE (4701)
and OFF-VALUE (4703) functions have the same sign or one of the values is
zero.
If the values have different signs, SYMMETRY cannot be selected and an
“ASSIGNMENT NOT POSSIBLE” message is displayed.
TIME CONSTANT
MODBUS register:
Relay output 1
Relay output 2
Data type:
Access:
4706
3811
4011
Float
read/write
Use this function to enter a time constant defining how the measuring signal
reacts to severely fluctuating measured variables, either very quickly (enter a low
time constant) or with damping (enter a high time constant).
Damping acts on the measuring signal before the switch status changes, and
consequently before switch-on or switch-off delay is activated.
The purpose of damping, therefore, is to prevent the relay output changing state
continuously in response to fluctuations in flow.
User input: fixed-point number: 0.00 to 100.00 s
Factory setting: 0.00 s
Function description
OUTPUTS RELAY OUTPUT (1 to 2) CONFIGURATION
0
-4
-10
4
10
Q
t
➀➁