Before plugging your tracer into the mains, check that the switch is set to OFF or 0.
If the switch no longer operates ON/OFF (0/I) you must stop using your tracer. Defective switches must be replaced by an approved
repairer.
Your system is a professional tool which must only be used by specialized, responsible operators. It should not be used by anyone else.
This machine is neither adapted nor intended to operate in an environment where there is a risk of explosion. Do not use this system in the
presence of liquids or flammable gases.
Make sure that the equipment is connected to a mains supply which complies with standard NFC 15-100. In addition, make sure that the
plug type is 3-pin 10/16 A (earthed) protected by a high sensitivity differential circuit-breaker, maximum 30 mA, and that the wall outlet is
easy to access.
LEDs used by the tracer:
Radiation emitted by the LEDs.
Do not look directly at the light when you use optical instruments.
Class 1M LED device
Maximum power emitted: 1.3 MW; 107 mJ
Wavelength: 400-800 nm
Name of the standard and publication date (IEC 60825-1:1993 / A2:2001).
7. ELECTROMAGNETIC WAVES
Electromagnetic waves emitted by television sets or radios, mobile phones, radio transmitters, etc. are liable to cause malfunctions in the
instrument. This instrument is also liable to interfere with radio or television sets, mobile phones, radio transmitters, etc. Avoid installing or
positioning devices which may create interference near the instrument. Likewise, never place the instrument on or near this type of device.
This could generate malfunctions or faults in the instrument.
Interference and micro-outages
The products have been designed to withstand interference and operate despite its presence and the possibility of micro-outages on the
network. However, if these malfunctions are too serious and abnormal, the machine cannot be guaranteed to operate normally.
Neksia > v1-02-14
124
USER MANUAL> GENERAL INFORMATION

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Essilor Neksia and is the answer not in the manual?

Essilor Neksia Specifications

General IconGeneral
BrandEssilor
ModelNeksia
CategoryEdger
LanguageEnglish

Related product manuals