DEPOLOX
®
Pool Installation / Commissioning
WT.050.570.000.DE.IM.0714 137
2.7
2.7.3 Connecting the Sample Water
CAUTION!
The sample water take-off must be located in front of the point of appli-
cation of the flocculant. This could otherwise influence measurement.
1 Sample water inlet with ball valve
2 Drain on the drain screw
3 Sample water outlet, only on pressurised version with ball valve
Fig. 41
Connecting the sample
water inlet
Never use copper tubing.
The pressure in the sample water inlet must always be within the
range 0.2 min. to 4 bar max. The pressure in the sample water inlet
must always be 0.2 bar higher than in the sample water outlet.
If the pressure is below 0.2 bar a pressure booster pump must
be used (see "Examples for sample water with pressure booster
pump" page 139 and page 140).
If the pressure exceeds 4 bar a pressure reducing valve must be
used.
To prevent long loop dead times, ensure that the pipes in the samp-
le water inlet are as short as possible.
An external strainer must be provided for the sample water inlet
(part no. W3T171391) with a mesh width of 0.5 mm.
1 2 3