DEPOLOX
®
Pool Replacing a Fuse or Battery
WT.050.570.000.DE.IM.0714 151
2.9
2.9 Replacing a Fuse or Battery
W
WARNING!
Only authorised and qualified electricians are permitted to open
the housing.
The device is not equipped with a mains switch.
Fuse Replacing a fuse
1. Disconnect the device from the power supply.
2. Remove the cover of the electronic module.
3. Remove all plug-in cards of the sensor measuring module
(2, Fig 48.
4. Remove the mA-OUT plug-in card (3).
5. Remove the relay card (4)
6. Remove the 4 outer screws on the metal cover (1).
7. Remove the metal cover.
8. Replace the plug-in fuses (5). TR5 1A slow-blow (W3T172031).
9. Reassemble the device.
Fig. 47 Replacing a fuse (5) or the battery (6)