DEPOLOX
®
Pool Operation
WT.050.570.000.DE.IM.0714 41
1.3
Calibration
(Menu 2.1)
Basic display Refer to main menu
Calibration Display of all available calibration options
Calibration Cl
2
(refer to section 1.3.4)
Cal. zero point with softkey "ENTER" the display is set to
"0.00 mg/l"
Cal. DPD within measuring range
Temporary Setp. refer to Calibration ORP
Calibration pH
(refer to section 1.3.4)
Cal. pH 7.00 6.85 ... 7.15 pH
Cal. pH within measuring range
Offset pH -1.00 ... +1.00 pH
Cal. at temp. 00 ... 50 °C
Calibration ORP
(refer to section 1.3.4)
Cal. mV within measuring range
Temporary Setp. Temporary setpoint until 2 hours after
ORP calibration
(only with CEDOX control)
Calibration Cl-N
Enter free Cl
2
within measuring range
Cal. total within measuring range
Cal. combined within measuring range
Calibr. conductivity
Cal. mS/cm within measuring range
Cal. °C 00.0 ... 50.0 °C
Calibr. temperature
Cal. °C 00.0 ... 50 °C