ST80/ST80L Flow Meter OPERATION
Fluid Components International LLC 71
Resetting the Totalizer Count
Repeat step 1 above, but specifying register #4117 instead (Length = 1).
Repeat step 2 above (no need to change serial configuration if already configured).
Double click on the register number (see pointer in Figure 55 below). A pop-up Write Register window displays. Enter the defined hex
value, 0xABCD, in the window’s value field, and then click Update.
Figure 55ModScan32 Connected to Modbus Device with Register 4117 on Display (Totalizer Reset)
Starting/Stopping the Totalizer Count
Repeat step 1 above, but specifying register #4118 instead (Length = 1).
Repeat step 2 above (no need to change serial configuration if already configured).
Double click on the register number (see pointer in the figure below). A pop-up Write Register window displays. Enter the defined
value (1 = start or 0 = stop) in the window’s value field, and then click Update.
Figure 56ModScan32 Connected to Modbus Device with Register 4118 on Display (Totalizer Start/Stop)