EasyManuals Logo

FCI ST80 Series User Manual

FCI ST80 Series
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #93 background imageLoading...
Page #93 background image
ST80/ST80L Flow Meter TROUBLESHOOTING
Fluid Components International LLC 85
Constant Power Configuration Troubleshooting
Equipment List
250 Ω 0.01% resistor
2 digital multimeters (DMM)
Delta R Calibration Data Sheet (serial number specific by instrument and group)
FES-200 flow element simulator
FES-200 interface cable for ST80/ST80L (022610-xx)
Alternative to FES-200:
2 ea. Precision Decade Resistance Box, 0.1% (1 kΩ large step, 0.01 Ω small step)
Delta R Check for Constant Power Configured Units
Note:
If the flow meter’s parameters have been changed, calibrations may be inaccurate or factory authorized changes
have been made. Consult a factory service representative.
Each flow meter configured at the factory for Constant Power is provided with a Delta R data sheet that lists the differential resistance
values that correlate to the flow meter’s calibration. Resistance substitution instruments like the FES-200 can be used to check instrument
calibration and verify correct operation of the flow transmitter using the Delta R data sheet.
To verify the transmitter is working properly, the sensor head must be disconnected and precision resistance (Delta R) values from the
FES-200 are substituted. Then by measuring the transmitter output and display it can be determined whether the transmitter is still within
factory specification.
Delta R Check
1. Verify the Delta R data sheet has the same serial number and group number as the flow meter calibration that is being verified.
2. Turn transmitter power OFF.
3. Disconnect a flow element sensor from the ST80/ST80L transmitter (TB1) and connect the FES-200 cable connector in its place. See
Figure 62. Precision decade boxes can be used in place of the FES-200. See Figure 63 for decade box wiring.
4. Connect a DMM to the transmitter 4-20 mA output by either method A or B as shown in Figure 61 below.
a. To read 1 to 5 volts, disconnect both output loop wires and connect a precision 250 Ω resistor across the output terminal. Then
connect the DMM, set to DC volts (V), across the resistor to read its voltage drop.
b. To read 4 to 20 mA current, disconnect the output loop and connect the DMM, set to milliamps (mA), in series with the output
circuit to read the current flow.
Figure 61DMM Hookup to Measure 4-20 mA Output
5. Turn transmitter power ON and allow the instrument 10 minutes to stabilize.
6. Verify the transmitter is in the calibration group matching the Delta R data sheet.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FCI ST80 Series and is the answer not in the manual?

FCI ST80 Series Specifications

General IconGeneral
BrandFCI
ModelST80 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals