20.14 Vial weight
Nominal weight for an empty vial. This is used to verify that the vial is empty to start
the pump.
i
Zero waste filling only. Refer to "Start Zero waste batch"
on page 91.
Requires a connected balance.
i
User must also set "Vial weight tolerance" below
i
It is recommended to use the average weight of multiple
vials or use the manufacturers data.
20.15 Vial weight tolerance
Sets the tolerance (+/-) of the nominal vial weight for a empty vial specified in "Vial
weight" above.
Always set vial weight tolerance value less than "Vial weight" above to prevent
accidental filling when no vial is present.
It is recommended to set vial weight tolerance less than "Volume" on page 1 if
possible to prevent accidental filling of a previously filled vial.
i
Zero waste filling only. Refer to "Start Zero waste batch"
on page 91.
Requires a connected balance.