Example A
• Target fill weight = 5 g
• Actual fill weight measured by the balance = 5.1175 g
• Upper auto recalibration tolerance limit = 7.5 g
• Lower auto recalibration tolerance limit = 2.5 g
While pump is dispensing, a filled vial is placed on the balance. The measured
weight falls within the auto recalibration tolerances. The pump automatically enters
calibration mode and performs a calibration.
Example B
• Target fill weight = 5 g
• Actual fill weight measured by the balance = 2.2010 g
• Upper auto recalibration tolerance limit = 7.5 g
• Lower auto recalibration tolerance limit = 2.5 g
While the pump is dispensing, a foreign object is placed on the balance. The
measured weight falls outside the auto recalibration tolerances. The pump will
continue dispensing.
23.3.3 Pausing a batch
To pause a batch press . Once the current fill has been completed, the
batch will pause.
When a batch is paused, the following options are available:
• Resume manual batch — This can be hidden, refer to "Operator access
restrictions" on page 105.
• Add note to report—"Add note to report" on page 89
• End batch—"End batch" on page 90
• Batch size—"Batch size" on page 81
• Fluid sample—"Fluid sample" on page 90
To resume filling press .
PF7/PF7+ Peristaltic Filler with QC14 Pumphead User Manual 87