78 PF7/PF7+ Peristaltic Filler with QC14 Pumphead User Manual
22.4 Multi-fill calibration without a balance connected
1. Place the container under the filling nozzle.
2. Select Multi-fill calibration.
3. Enter the number of calibration fills.
4. Start the calibration fills depending on the selected filling method:
Time delay between fills - Press or use the external input to start the
first calibration fill. Any further fills will start automatically after the Between fill
delay has elapsed.
Keypad to start each fill - Press to start each fill.
External input start each fill - Press or use the external input to start
each fill.