88 PF7/PF7+ Peristaltic Filler with QC14 Pumphead User Manual
23.3.4 Calibration whilst paused
1.
To pause a batch press . Once the current fill has been completed, the
batch will pause.
2. If using a balance, tare the balance using the vial to be filled.
3.
Press to access Calibrate mode.
When batch is paused calibrate mode is used to:
l Enter the calibration value
i.
No connected balance— Use the numeric keypad. Press to
confirm.
ii. Connected balance— Place the filled vial on the balance. The value
automatically displays. Press to confirm.
l Dispense a calibration fill dispense a calibration fill before the calibration
value is entered.
i
A rolling average of previous fill weights can be used for
calibration. Refer to "Calibration settings" on page 112.