Calibration value entry
5. Enter the dispensed value using the numeric keypad. The Target weight is the
total value of all fills.
6. Press .
Calibration outside an acceptable level
If the calibration value entered is outside of ±95% of the target value:
• A warning screen is displayed.
• Calibration cannot continue.
• Check the value was entered correctly.
• Check the tube size is as per the recipe.
• Check the fluid path is set up correctly.
Calibration with fill tolerance limits
Refer to "Fill tolerance" on page 63 to enable fill tolerance limits.
PF7/PF7+ Peristaltic Filler with QC14 Pumphead User Manual 79