5322A
Operators Manual
134
[SOUR]:SAF:LIN:CLE
Description: Clears the maximum (peak) value of current flowing through the
selected resistance. This command switches the Product into the
Line Impedance function mode.
Example: SAF:LIN:CLE
[SOUR]:SAF:LIN[:CURR]:PFC?
Description: This command returns the value of the prospective fault current.
The Product returns the current value using exponential format.
Example: 100 A is returned as 100e+00. If not already selected,
this command switches the Product into the Line Impedance
function mode.
Query: SAF:LIN:PFC?
[SOUR]:SAF:LIN:MAN(?) <DNPD>
Description: This command sets the manually-entered value of residual
impedance. This command switches the Product into the Line
Impedance function mode.
Parameters: <DNPD> Parameter represents the value of residual
impedance in Ω. See the Specifications
(found online at www.Flukecal.com) for
acceptable ranges.
Example: SAF:LIN:MAN 0.87
Query: SAF:LIN:MAN? The Product returns the manual entered
value of residual impedance using
exponential format. Example: 0.72 Ω is
returned as 0.720e+00.
[SOUR]:SAF:LIN:PONS(?) <CPD> { ON | OFF }
Description: This command sets the Power ON (Start Up) switch. This
command switches the Product into the Line Impedance function
mode.
The state of the Power ON switch is the same for the Loop and
Line functions.
Parameters: <CPD> Parameter represents the switch state.
GROund grounds the Low-Resistance
output
FLOat floats the Low-Resistance output
from ground
Query: SAF:LIN:PONS? The Product returns the set switch state.