Multifunction Electrical Tester Calibrator
Supported SCPI Commands
135
[SOUR]:SAF:LIN:SRES(?) <DNPD>
Description: This command sets the series resistance of the connection cables.
This command switches the Product into the Line Impedance
function mode.
Parameters: <DNPD> Parameter represents the value of series
resistance in Ω. See the Specifications
(found online at www.Flukecal.com) for the
acceptable ranges.
Query: SAF:LIN:SRES? The Product returns the value of series
resistance using exponential format.
Example: 0.72 Ω is returned as 0.720e+00.
[SOUR]:SAF:LIN:RESC
Description: Starts internal measurement of power line residual impedance. If
not already selected, this command switches the Product into the
Line Impedance function mode.
Example: SAF:LIN:RESC
[SOUR]:SAF:LOOP[:LEV](?) [<DNPD>]
Description: This command switches the Product into the Loop Impedance
function mode and sets the resistance value.
Parameters: <DNPD> Parameter represents the value of the
resistance expressed in Ohms. The Product
selects the nearest resistance value.
Example: SAF:LOOP 0.1 Loop Impedance 100 mΩ
Query: SAF:LOOP? The Product returns the set value of
resistance using exponential format.
Example: 900 Ω is returned as 900.0e+00.