5322A
Operators Manual
144
[SOUR]:SAF:HIPT:FUNC(?) <CPD> {DC | AC}
Description: This command selects dc or ac hipot timer mode. This command
switches the Product into the Hipot timer function mode.
Parameters: <CPD> DC sets to the dc hipot timer mode.
AC sets to the ac hipot timer mode.
Example: SAF:HIPT:FUNC AC Set to the AC hipot timer mode.
Query: SAF:HIPT:FUNC? The Product returns ac or dc.
[SOUR]:SAF:HIPT:PROB(?) <CPD> { OFF | 5KV | 10KV | 40KV }
Description: This command selects the internal or external high voltage probe
for hipot timer. If not already selected, this command switches the
Product into the Hipot timer mode.
Parameters: <CPD> OFF – hipot timer with no probe (V terminal)
5KV – hipot timer with the HV terminal
10KV – hipot timer with an external 10kV probe
40KV – hipot timer with an external 40kV probe
Example: SAF:HIPT:PROB 10KV Set to the external 10kV high voltage
probe in hipot timer mode.
Query: SAF:HIPT:PROB? Returns whether the multimeter is measuring
with a probe and if so, which probe is being used.
[SOUR]:SAF:HIPT:THR(?) <DNPD>
Description: This command sets the threshold level in Hipot timer mode. If not
already selected, this command switches the Product to Hipot
timer.
Parameters: <DNPD> Represents the value of the threshold level expressed in
percent.
Example: SAF:HIPT:THR 50 Set the threshold to 50%.
Query: SAF:HIPT:THR? The Product returns the threshold value in integer
format.
[SOUR]:SAF:HIPT:TIME?
Description: This command returns the measured time using exponential format.
Example: 2.15 s is returned as 2.150e+00. This command switches
the Product into the Hipot timer function mode.
Query: SAF:HIPT:TIME?