Multifunction Electrical Tester Calibrator
Supported SCPI Commands
143
[SOUR]:SAF:HIPL:VOLT:MAX?
Description: This command returns the maximum (peak) value of measured
voltage. Maximum value can be cleared using the command
SAF:HIPL:CLEar.
The Product returns the current value using exponential format.
Example: 230 V is returned as 230.0e+00. This command
switches the Product into the Hipot leakage current function mode.
Query: SAF:HIPL:VOLT:MAX?
[SOUR]:SAF:HIPL:CLE
Description: Clears the maximum (peak) measured voltage. This command
switches the Product into the Hipot leakage current function mode.
Example: SAF:HIPL:CLE
[SOUR]:SAF:HIPL:VOLT:THD?
Description: This command returns the THD value of measured voltage.
The Product returns the THD value using exponential format.
Example: 0% is returned as 0.00e+00. This command switches
the Product into the Hipot leakage current function mode.
Query: SAF:HIPL:VOLT:THD?
[SOUR]:SAF:HIPL:VOLT:RIPA?
Description: This command returns the Ripple absolute measured
voltage using exponential format. Example: 1 V is returned as
1.000e+00. This command switches the Product into the Hipot
leakage current function mode.
Query: SAF:HIPL:VOLT:RIPA?
[SOUR]:SAF:HIPL:VOLT:RIPR?
Description: This command returns the Ripple relative measured
voltage using exponential format. Example: 0 % is returned as
0.00e+00. This command switches the Product into the Hipot
leakage current function mode.
Query: SAF:HIPL:VOLT:RIPR?