All deviation diagrams show the actual number of active and inactive points, the current stand-
ard deviation (SD), and the total change in standard deviation (delta SD) when deviant points
are deactivated.
What is the effect of deactivating individual points?
Individual points are deactivated in order to lower the standard deviation value. Change in
standard deviation impacts the V and H results displayed in the results repeatabilty table. Res-
ults with a green check mark indicate results with better standard deviation.
ROTALIGNtouch EX 95
On-board help