Command 206 (0xCE): Write Fluid Information
This command is to write fluid information.
Table 134: Request Data Bytes for Write Fluid Information
Byte Format Description
0 - 3 Unsigned-32 Fluid Type:
0: Other
1. Water
4 - 7 Float Fluid SOS
8 – 11 Float Fluid minimum SOS
12 - 15 Float Fluid Maximum SOS
16 - 19 Float Fluid Temperature
Table 135: Response Data Bytes for Write Fluid Information
Byte Format Description
0 - 3 Unsigned-32 Fluid Type:
0: Other
1. Water
4 - 7 Float Fluid SOS
8 – 11 Float Fluid minimum SOS
12 - 15 Float Fluid Maximum SOS
16 - 19 Float Fluid Temperature
Table 136: Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1Undefined
2 Error Invalid Selection
3-4 Undefined
5Error Too Few Data Bytes Received
6 Error Device-Specific Command Error
7Error In Write Protect Mode
8-15 Undefined
16 Error Access Restricted
17-127 Undefined