EasyManua.ls Logo

GE Giraffe OmniBed Carestation CS1 - 5.5 Status

GE Giraffe OmniBed Carestation CS1
308 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5: Troubleshooting
Status
5-22 Service Manual
5.4.4.4 Sensor Housing Temperature Sensor Temperature to Resistance Curve
5.5 Status
Access the Status screen from the main service menu. See “Accessing Service Mode” on page 4-1.
The Status screen displays status of all self tests the software runs continuously. These include: incubator
heater on (INCHTRON), warmer heater on (WRMHTRON), incubator/warmer heater off (I/WHTROFF),
humidifier heater on (RHHTR), remote monitoring data stream (RS-232LOOP), incubator fan on (FANON), and
incubator fan off (FANOFF). The current software revision of the options installed on this specific device
(Humidifier, Scale, SPO2, etc.) is shown.
If the RS-232 is not installed RS-232LOOP will display N/A. To test the circuit if the RS-232 is working, connect
the RS-232 to a working RS-232 compatible device or short pins 2 and 3 on the 9 pin connector.
Temperature Resistance
15 °C 15.720 kOhm
25 °C 10.000 kOhm
30 °C 8.056 kOhm
35 °C 6.530 kOhm
40 °C 5.326 kOhm
45 °C 4.369 kOhm
50 °C 3.604 kOhm
60 °C 2.491 kOhm
FIGURE 5-2. Status Screen

Table of Contents

Related product manuals