EasyManua.ls Logo

GE Giraffe OmniBed Carestation CS1 - 7.13 Control Panel and Display Module Procedures

GE Giraffe OmniBed Carestation CS1
308 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 7: Repair Procedures
Control Panel and Display Module Procedures
7-56 Service Manual
7.13 Control Panel and Display Module Procedures
7.13.1 Probe Panel
Refer to “Figure 7-52 Probe Panel” and “Figure 8-1 Probe Panel Assembly”.
1. Using a 2.5 mm hex key, loosen the 2 screws in the keyhole slots and remove the 6 remaining screws
that secure the control panel cover, then remove the cover. Disconnect the wire harnesses coming from
the probe panel.
2. Remove the 2 screws at the bottom corners that hold the probe panel to the enclosure.
3. Remove the probe panel assembly.
4. While tilting the panel enclosure forward, use a 3 mm hex key to remove the 4 screws that hold the
panel frame in place. Add new connectors through the panel frame as appropriate.
Note: The power switch panel must be on the far left. The position of the remaining panels is not critical.
FIGURE 7-52. Probe Panel
Power switch
Probe panel housing
Electrical enclosure
Probe panel housing
PF Battery holder
PF Battery

Table of Contents

Related product manuals