Chapter 6: Alarms and System Failure Codes
Alarms
Service Manual 6-3
6.1.6 Bed Height Pedals Disabled
6.1.7 Bed Raise Pedal Pressed
6.1.8 Bed Lower Pedal Pressed
Cause or Circumstances Failure Action
The Bed Height Pedals Disabled
alarm activates if the bed raise
or lower pedals have been
disabled on either the Service
screen or the Setup screen, and
one of the pedals is pressed.
Bed pedal is disabled in service
Mode.
Bed pedal is disabled in Setup
screen.
Release the pedal.
If the canopy pedal is disabled in
the service screen and you wish to
use it, contact authorized service
personnel to reactivate it on the
service screen.
Cause or Circumstances Failure Action
The Raise Pedal Pressed alarm
activates if the pedals are
pressed while turning on the
device.
Pedals are pressed while
turning on the device.
Release the pedal.
Cause or Circumstances Failure Action
The Lower Pedal Pressed alarm
activates if the pedals are
pressed while turning on the
device.
Pedals are pressed while
turning on the device.
Release the pedal.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe OmniBed Carestation CS1 and is the answer not in the manual?

GE Giraffe OmniBed Carestation CS1 Specifications

General IconGeneral
BrandGE
ModelGiraffe OmniBed Carestation CS1
CategoryMedical Equipment
LanguageEnglish

Related product manuals