Chapter 6: Alarms and System Failure Codes
System Failure Codes
Service Manual 6-25
6.2.37 System Failure 63
6.2.38 System Failure 70
6.2.39 System Failure 71
6.2.40 System Failure 81
6.2.41 System Failure 82
Cause(s) Action(s)
Control board has
incompatible software
version.
Replace control board software with latest version
Cause(s) Action(s)
Unexpected SBC software
problem
Power cycle the device. If issue persists replace SBC.
Cause(s) Action(s)
Unexpected SBC software
problem
Power cycle the device. If issue persists replace SBC.
Cause(s) Action(s)
Unexpected SBC software
problem
Power cycle the device. If issue persists replace SBC.
Cause(s) Action(s)
Unexpected SBC software
problem
Power cycle the device. If issue persists replace SBC.