Chapter 4: Understanding the Service Interface
Calibration
Service Manual 4-5
3. Wait for the prompt to disappear before powering off unit.
4.6 Calibration
The Calibration screen allows to view status and calibrate Oxygen, humidity, scale, and line voltage.
4.7 Set Time
The Set Time screen allows you to set the real time clock for time, day and date.
Pulldown Item Description
Language Selects the language to display on-screen in the user interface. English is the
default language.
Temp Units Select the temperature units displayed. ºC, ºF, or ºC only are available. ºC only
prevents the Fahrenheit option from showing on the user Setup screen (Celsius
is factory set default).
Volume Select one of four alarm volume settings. 1 is minimum; 4 is maximum.
Skin Temp Alarm
(Baby Mode)
Set the default Hot patient/Cold patient alarm to activate when either 0.5ºC or
1.0ºC difference is read between a set temperature and the patient
temperature.
Elevate bed Disable or enable the bed Raise/lower pedal switches. If the pedals are disabled
on the service screen, they cannot be enabled on the user Setup screen.
Canopy Disable or enable the canopy foot pedal switches, they cannot be enabled on
the user Setup screen.
Note: The hand switches remain active when the Canopy is disabled on the
service screen.
Scale Unit Select from Grams, Pounds, or Grams Only. Grams Only prevents the pounds
option from showing on the user Scale screen. Grams is the factory default.
Scale Resolution Select from available scale resolution settings: 10, 5, 2, or 1 gram (10 grams is
the factory default).
Skin temp alarm (Air/
Manual Mode)
Use to enable or disable the skin temperature alarms in the Manual Mode. If
disabled on the service screen, it cannot be enabled on the user setup screen.
The alarm will not activate if a patient set point has not been entered.
Hands free alarm Select to enable or disable the Hands Free Alarm Silence.