Appendix B: Specifications
B-4 Service Manual
Humidity
Servo Control Accuracy* ± 10% for settings up to 85%; minimum
75% for settings > 85%
Humidity control setting vs
average humidity at 10 cm above
center of bed.
Ramp-up Time ≤ 50 minutes Time to reach 75% RH with a 39°C
control temp from cold start in
25°C, 50% RH room.
Note: The above item noted with an asterisk (*) indicates essential performance.
Operating Time without Refill > 12 hours Operational time at 65% RH
control setting with one filling of
reservoir in 25°C, 50% RH room.
Mechanical
Height Canopy closed, bed lowered 152 cm
Canopy closed, bed raised 178 cm
Canopy open, bed lowered 209 cm
Canopy open, bed raised 239 cm
Width 66 cm
Depth 114 cm
Weight 149 kg
Mattress Size 48.8 cm X 64.8 cm
Bed Tilt ± 12 degrees continuous tilt
The Maximum Patient Weight 40 kg (88 lbs)
Option Load Limits
Storage Drawer Package 7 kg
Instrument Shelf 9 kg
Monitor Shelf 22 kg
Rotating I.V. Pole 9 kg
Total Each Mounting Rail 22 kg
Uninterruptable Power Supply (UPS) (Not available with 230V devices)
For UPS specifications, refer to the Addendum: UPS for Giraffe and Panda Products (provided with the UPS).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe OmniBed Carestation CS1 and is the answer not in the manual?

GE Giraffe OmniBed Carestation CS1 Specifications

General IconGeneral
BrandGE
ModelGiraffe OmniBed Carestation CS1
CategoryMedical Equipment
LanguageEnglish

Related product manuals