Measurement and Analysis
120 Vivid i User’s Manual
2378958-100 Rev. 02
See the Status bar
to get the next step
to perform.
6. Press SET to anchor the start point of the measurement.
7. Trackball the cursor to draw the length.
8. Press
SET to anchor the second point.
9. Drag the cursor with the Trackball to outline the area of
interest.
The measurement
display color on the
2D image changes
from green to red af-
ter completion of
the measurement.
The current area, circumference and Area/Length Volume
(Va-l) values are displayed in the Measurement result table
(see Figure 6-1) and are instantaneously updated when
moving the cursor.
10. Press
SET to complete the measurement.
The measurement results including Vmod (Simpson) are
displayed in the Measurement result table (see
Figure 6-1).
The measurements
displayed on the 2D
image and the cor-
responding results
are numbered.
11. To assign a label to the measurement, see page 115.
12. Repeat steps 5 through 10 to make additional volume
measurements.
2D Depth measurements
The measurements described in this section enable depth
measurement from the probe to a selected point.
To perform a depth measurement:
1. Generate the 2D image.
2. Press
FREEZE to stop the cineloop.
3. Press
MEASURE on the Control Panel.
See the Status bar
to get the next step
to perform.
4. Press the assignable POINT to select the depth
measurement function.
5. Trackball the cursor to the position to measure.
The current distance from the probe is displayed in the
Measurement result table and is instantaneously updated
when moving the cursor.
The measurements
displayed on the 2D
image and the cor-
responding results
are numbered.
6. Press SET to anchor the point.
The depth value (cm) is displayed in the Measurement
result table.