Archiving
190 Vivid i User’s Manual
2378958-100 Rev. 02
A stand-alone scanner and a
stand-alone EchoPAC PC environment
In this scenario the EchoPAC PC (one or several) is used for
review of studies acquired on one or more Vivid i without being
connected via a private or a local area network.
Images can be stored on the scanners internal harddrive
(recommended) or on a dedicated MOD.
Images stored on the internal harddrive
Figure 8-29: A stand alone scanner and a stand alone EchoPAC PC
environment with images stored on the scanner’s harddrive
In this configuration images are first stored on the scanner’s
harddrive and then exported from the scanner’s harddrive to a
MOD and finally imported from the MOD to the EchoPAC PC’s
internal harddrive.
Data management
Scanners dataflow configuration
1. Select the LocalArchive-Int.HD dataflow as default
dataflow.
The local database is used for patient archiving. Images
are stored to internal harddrive.
Export from Vivid i
1. Export the data (images, demographics, measurements
and report) for the patient(s) to be reviewed on a blank
Vivid i: dataflow LocalArchive-IntHD
EXP: export from LocalArchive-Int.HD to Removable MOD Archive
IMP: import from Removable MOD Archive to LocalArchive-Int.HD
EchoPAC PC: dataflow LocalArchive-Int.HD