Measurement and Analysis
Vivid i User’s Manual 117
2378958-100 Rev. 02
Figure 6-5: The Enter new parameter window
4. Type a name for the parameter label.
5. Press OK.
The user defined parameter label is assigned to the
selected measurement.
Cardiac measurements
2D Measurements
2D Length measurements
1. Generate the 2D image.
2. Press
FREEZE to stop the cineloop.
3. Press
MEASURE on the Control Panel.
4. Select Caliper in the Measurement Menu (see Figure 6-1).
5. Trackball the cursor to the start point of the measurement.
6. Press
SET to anchor the start point of the measurement.
See the Status bar
to get the next step
to perform.
7. Trackball the cursor to the measurement end point.
The current distance value is displayed in the
Measurement result table and is instantaneously updated
when moving the cursor.
The measurement
display color on the
2D image changes
from green to red af-
ter completion of
the measurement.
8. Press SET to anchor the end point of the measurement.
The measurement result is displayed in the Measurement
result table.
9. To assign a label to the measurement, see page 115.
10. Repeat steps 5 through 8 to make additional length
measurements.