Archiving
216 Vivid i User’s Manual
2378958-100 Rev. 02
Importing patient records/examinations
1. Insert the removable media of the source archive in the
corresponding drive (MO drive or CD-ROM).
2. Press
PATIENT on the Front panel, then select Patient List.
The Search/Create Patient window is displayed
(Figure 8-12, page 169).
3. Select destination archive in the Dataflow field:
LocalArchive-Int.HD: imports data to the local archive.
RemoteArch-RemoteHD: imports data to an Echo
server (network) or an EchoPAC PC (direct connect).
4. Press Import in the Search/Create Patient window.
The Import dialogue window is displayed (see
Figure 8-47).
Figure 8-47: The Import Dialogue window
5. Select one of the following source archive from the Source
drop-down menu:
Removable MOD Archive: imports raw and DICOM
data (if present) from a MOD.
Pure DICOM MOD525: imports DICOM data only from
a MOD.
Removable CD Archive: imports raw and DICOM data
(if present) from a CD-R.
Pure DICOM CD-R: imports DICOM data only from a
CD-R.
Remote Import/Export Archive: imports raw and
DICOM (if present) data from an Echo server (network)
or EchoPAC PC (direct connect or network).
Query Retrieve: imports data from a DICOM server.
6. Press OK.