Archiving
232 Vivid i User’s Manual
2378958-100 Rev. 02
DICOM spooler
DICOM spooler displays the current DICOM output jobs. The
jobs may be Storage, Print, Modality Performed Procedure
Step or Storage Commitment. The DICOM spooler is used for
checking the current job's status when a job is saved or when
the total spooler status on the right of the Archive windows
displays an error.
From the DICOM spooler the user can also:
• Delete non-active jobs
• Resend a job that has failed or is in hold
• Send a job that has failed or is in hold, to a new destination.
The job's status displayed in the DICOM spooler window can
be:
• Pending: the job is complete, waiting to be active.
• In hold: the job is incomplete, waiting for more images.
• Append: the job is connected to the destination device but
is waiting for more images (Direct store function).
• Active: the job is complete and connected to the
destination device.
• Failed: the job is complete but one or more images failed to
transmit to the destination device.
• Done: the job is saved to the destination device. The jobs
that are done are removed from the spooler after a while.
Starting the DICOM spooler
1. On the alphanumeric keyboard, press and hold down the
ALT key and press S.
The DICOM spooler window is displayed (see
Figure 8-66).
The DICOM spooler window is automatically updated every
minute. Press Refresh to update the information displayed at
any time.

Table of Contents

Other manuals for GE Vivid i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid i and is the answer not in the manual?

GE Vivid i Specifications

General IconGeneral
TypePortable Ultrasound System
Transducer Ports2 active ports
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Imaging Modes2D, M-Mode
ConnectivityDICOM, Ethernet, USB
ApplicationsVascular, Abdominal
WeightApproximately 5.5 kg (12.1 lbs)
Battery LifeUp to 1 hour (depending on usage)
ProbesPhased array, Linear array, Convex array