User maintenance
Vivid i User’s Manual 337
2378958-100 Rev. 02
Virus Protection
To minimize virus vulnerability Vivid i is configured with a
minimal set of open ports and with all network services not
actively used by the system closed down. This significantly
reduces the risk of a virus attack on Vivid i. GE is continuously
judging the need for additional actions to reduce vulnerability of
equipment, this includes vulnerability scanning of our products
and evaluation of new security patches for the 3rd party
technology used. Microsoft (and other) security patches that
addresses serious issues with Vivid i will be made available to
customers after GE verification of those patches.
Cleaning the unit
General Cleaning
Frequent and diligent cleaning of the Vivid i ultrasound unit
reduces the risk of spreading infection from person to person,
and also helps to maintain a clean working environment.
LCD Display Cover
On a weekly basis, moisten a soft, non-abrasive folded cloth or
sponge with a mild, general purpose, non-abrasive soap and
water solution. Do not use any solution containing abrasive
powder or strong chemicals such as acid or alkaline.
Squeeze excess liquid from the cloth/sponge, them wipe down
the top, front, back and both sides of the unit. Do not spray
any liquid directly onto the unit!
1. Rinse the cloth/sponge with clean running water and wipe
the unit surfaces again.
2. Use a dry, soft, lint-free cloth to dry the unit surfaces.
CAUTION
When performing Cleaning Procedures, to prevent the risk of
system damage, always observe the following precautions:
Use only cleaning materials and solutions recommended in the
procedures described in Table xx on page xx.
Do not use any solutions or products not listed in the Vivid i User
Manual.
Do not spray any liquid directly onto the Vivid i covers, LCD Display
or keyboard!
Do not allow any liquid to drip or seep into the system.
Prior to cleaning, turn OFF power to the system.