Getting started
46 Vivid i User’s Manual
2378958-100 Rev. 02
Connectivity on the Vivid i ultrasound unit
Refer to page 186
for further informa-
tion on connectivity
setup.
The connectivity on the Vivid i Ultrasound unit is based on the
Dataflow concept. A Dataflow is a set of pre-configured
services (e.g. DICOM services like storage, worklist, verify etc.
or other service types like video print, standard print or
messaging). When starting an examination, the user selects a
pre-configured Dataflow that will automatically customize the
ultrasound unit to work according to the services associated to
the Dataflow.
Selecting a Probe and an Application
The combination
Probe-Application
may be user-de-
fined. See page 291
for information on
probe/application
configuration.
Probes and their related applications are selected from the
Probes and applications pop-up menus as described below.
Only probes currently connected are displayed in the pop-up
menu. Only applications appropriate for the type of probe
selected are shown.
To select a probe and an application
1. Press APPLICATION on the control panel.
A list of the connected probes is displayed.
2. Trackball to the desired probe.
3. Press
SET.
An Application menu for the selected probe is displayed.
4. Trackball to the desired application.
To select a probe
with the default ap-
plication, press SET
twice on the actual
probe.
5. Press SET to launch the application.
CAUTION
Make sure that the probe and application names displayed on the
screen correspond to the actual probe and application selection.
Check that the correct TI category is displayed (see Chapter 14,
"Thermal Index" on page 353). TIB must be displayed when a
fetal application is selected.