Probes
272 Vivid i User’s Manual
2378958-100 Rev. 02
Inspecting the probe
After each use:
1. Inspect the lens, the probe housing and the cable
(Figure 10-3.
2. Look for damage that might allow liquid into the probe.
Before each use:
1. Inspect the lens, the probe housing and the cable
(Figure 10-3).
2. Look for damage that might allow liquid into the probe.
3. Test the functionality of the probe.
Figure 10-3: Probe parts
CAUTION
If any damage is found, DO NOT use the probe until it has been
inspected and released for further use by a GE service
representative.
1. Housing
2. Strain relief
3. Seal
4. Lens