Measurement and Analysis
Vivid i User’s Manual 123
2378958-100 Rev. 02
8. Press SET. The IVSd measurement end point is anchored
and the value is displayed in the Measurement result table.
The end point of the IVSd is also the start point for the LVIDd.
1. Trackball to the end point of the LVIDd measurement.
2. Press
SET. The LVIDd measurement end point is anchored
and the value is displayed in the Measurement result table.
The end point of the LVIDd is also the start point for the
LVPWd.
1. Trackball to the end point of the LVPWd measurement.
2. Press
SET. The LVPWd measurement end point is
anchored and the value is displayed in the Measurement
result table.
3. Repeat steps 5, through 2 to measure IVS, LVID and LVPW
in systole.
RV study
The RV study consists of measurement in fixed-time mode of
Right ventricular internal dimension (RVID) in both diastole and
systole.
To perform RV study
1. Generate the M-Mode image.
2. Press
FREEZE to stop the cineloop.
3. Press
MEASURE on the Control Panel.
4. Select RV study in the Measurement Menu.
5. Trackball the cursor along the time axis to the required point
to start measurement of RVIDd.
6. Press
SET. The starting point for the measurement is
anchored.
The current value is
updated while mov-
ing the cursor.
7. Trackball to the end point of the measurement.
8. Press
SET. The measurement end point is anchored and
the RVIDs measurement value is displayed in the
Measurement result table.
A new free-moving cursor is displayed on the image, ready
for the next measurement.
9. Repeat steps 5, through 8 to measure RVIDs.
The RVIDs value is displayed in the Measurement result
table.