Archiving
194 Vivid i User’s Manual
2378958-100 Rev. 02
4. In the Remote archive setup area enter:
• The IP address for the remote archive
• The name of the remote archive
5. Press Save settings and reboot the system.
Direct connect with online acquisition
Data management
Scanner’s dataflow configuration
1. Select the RemoteArch-RemoteHD dataflow as default
dataflow.
When saving the study on the scanner, the images are
transferred from the scanner’s image buffer to the
harddrive of the EchoPAC PC. Patient demographics,
measurements and reports are transferred on the fly when
entering the information on the Vivid i.
Review on the EchoPAC PC workstation
1. Select the LocalArchive-Int.HD dataflow on the
EchoPAC PC (can be configured as default dataflow).
2. Press Archive and select the patient to be reviewed.
CAUTION
Do NOT open a study on the EchoPAC PC workstation before the
study is closed on the scanner.