If only the O
2
backup gas cylinder is used, the
Air and/or N
2
O backup gas cylinder pressure
check can be disabled in Service and Settings
– Startup configuration, see Chapter 9, page
178.
Circuit compliance compensation is calculated
during the leakage tests and presented with
the test results.
If N
2
O central gas supply pressure is
connected while the N
2
O function is
deactivated, the system checkout will fail.
WARNING! If the System checkout is
unsuccessful, the system must not be
connected to the patient until the
malfunction is corrected. If the malfunction
cannot be corrected by the operator, the
system must be turned Off, taken out of
operation and serviced by personnel trained
and authorized by MAQUET.
4.2.5 Finalization
01-01 13 00
1 2 3
Once all checks are completed, a summary is
displayed on the screen. Press the Standby
touch pad located at the bottom right of the
screen to enter Standby mode.
Checks can be repeated by pressing any of
the touch pad buttons located in the bottom
left area of the panel screen:
1. Redo System checkout
2. Redo leakage check
3. Redo vaporizer check
FLOW-i 4.2, User's Manual
57
| Startup and system checkout |
4 |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals