Use
Daily inspections before use
4
Maquet PowerLED II
IFU 01811 EN 10
45 / 116
Fig.41: Lighthead gaskets
Lighthead gaskets
1. Check that the lighthead seals are in the
proper position and in good condition.
2. If a problem is noted, contact technical
support.
Fig.42: Lighthead underside
Lighthead underside
1. Check that the underside is not damaged.
2. If a problem is noted, contact technical
support.
Fig.43: Condition of lighthead keypad
Lighthead control keypad
1. Check that the lighthead control keypad is
in good condition and in the proper posi-
tion.
2. Press the ON/OFF button for 5 seconds.
Ø All buttons and warning indicators are
backlit.
3. If a problem is noted, contact technical
support.